The infant incubator from a hygienic and HTO perspective : Using ATP luminescence to identify problem areas and suggesting solutions

University essay from KTH/Skolan för teknik och hälsa (STH)

Abstract: Healthcare associated infections (HCAI) are a major problem in healthcare today. Preterm infants have problems keeping their body temperature within normal boundaries due to heat-loss. They therefore need special care that is administered with the help of incubators, which help minimise the heat loss via convection. Within neonatology the incubator has been identified as one of the contributing factors to HCAI due to the warm and humid environment, making it easy to spread nosocomial flora. To assess if the incubator is a factor in the spreading of HCAI this project has focused on ATP+AMP (total ATP) luminescence measurements to find areas in the incubator that are likely to contribute to the spread of HCAI and suggesting solutions to some of these. Adenosine triphosphate (ATP) is found in both organic debris and bacteria and is therefore a good indicator of a problem area due to organic debris acting as nutrients. Only the incubator box of the Giraffe® OmniBed® incubator was studied. The cleaning process was observed on multiple occasions and together with interviews resulted in a number of 29 hypothesised problem areas, on or within the incubator box, that were measured before and after cleaning. The results show that incubators collect a substantial amount of total ATP during its use. Measurements also show that parts that are cleaned by a disinfector are cleaner than those parts that are cleaned manually. Areas on the main compartment became more contaminated after cleaning which further indicated that the design of the incubator needs improving. It was also concluded that there often was residue from soap left on the surface of the main compartment resulting in inhibition of the total ATP luminescence reaction. This resulted in unrealistic low values due to the inclusion of foam and soap in the sample and as a result 45 out of 570 measurements were excluded. Caution is advised when using the Kikkoman total ATP luminescence method, especially on the main body (chassis) of the incubator. A steam vapour cleaner and flask cleanser brush could be used to better reach and clean areas such as cavities and around the bed heating element. The use of the steam vapour cleaner could also diminish or eliminate the use of surface disinfectant that causes red irritated eyes and dizziness in the cleaning staff. The functionality of the incubator was found to be excellent but major improvements can be made in the incubator design to make the cleaning of the incubator easier.

  AT THIS PAGE YOU CAN DOWNLOAD THE WHOLE ESSAY. (follow the link to the next page)