Evaluation of the reconstruction algorithm, Ordered Subsets Expectation Maximization, in whole body Positron Emission Tomography

University essay from Institutionen för medicinsk teknik

Abstract: A Positive Electron Tomography/Computed Tomography devise was installed in theX-ray section at US Linköping in May 2007. Positive Electron Tomography examinations with 18F-fluoro-deoxy-glucose are mainly used for tumor examinations. During 2007 occurred approximately 200 examinations and in 2008 600 are planned. Today there are two reconstruction methods commercially available, Filtered Back projection and Maximum Likelihood Expectation Maximiza tion, used in the faster version called Ordered Subsets Expectation Maximization. The image quality in Positive Electron Tomography depends on the choice of reconstruction method and the settings of its parameters. We have performed a physical phantom study with Positive Electron Tomography to determine optimal parameters for the iterative reconstruction algorithm Ordered Subsets Expectation Maximization. To find out whether or not the quality of the image can be improved, so that the patient received radiation dose and/or examination time can be lowered. The phantom used was a NEMA IEC Body PhantomTM, designed to mimic smallhot lesions typical in 18F, Fluorine-18 PET, and all calculations were done according to the NEMA NU2-2001 protocol. The main conclusion from this project is that a higher level of contrast can be reached, compared to the one clinically obtained today. Using more iterations then recommended from the manufacturer.